Trace Oxygen Analyzer

Operation 4

4-19

Teledyne Analytical Instruments

4.13 The

Analysis Mode

This is the normal operating mode of the analyzer. In this mode the

analyzer is monitoring the sample, measuring and displaying the amount of
oxygen, and reporting alarm conditions.

Normally, all of the functions automatically switch back to the Analysis

Mode ANALYZE screen when they have completed their assigned opera-
tions. After four or five minutes in the MAIN MENU without any action by
the operator, the analyzer automatically switches itself back to the ANA-
LYZE screen. ESCAPE, asserted one or more times, depending on the
starting point, also switches the analyzer back to the ANALYZE screen.

IMPORTANT: In the event of loss of flow through the analyzer, if the

vent is vented to a location of high oxygen content,
oxygen will back diffuse through the vent line and in
most cases quickly saturate the cell with oxygen which

can then require a quite long purge down time for the

sensor when then exposed to low oxygen concentra-
tions. In the event that flow is to be interrupted into the
analyzer, it is suggested that the user do one of the
following:

1.

Bag the sensor in nitrogen during this time

2.

Install a shut off valve on the vent port of the ana-
lyzer or somewhere within the users sample system.