Toshiba Sanitary Electromagnetic Flowmeter TIC-LF494B User Manual

Page 3

Advertising
background image

TIC-LF494B

3

Model LF610 and LF612 converters

Input signals

Analog signal — the voltage signal from detector,

proportional to process flow rate (for LF612

separate type converter).

Digital input DI (opt.)

Signal type: 20 to 30Vdc voltage signal

Input resistance: 2.7k

Number of inputs: one point

DI function — One of the following functions

can be assigned to the optional DI signal.

Range switching — Selects either the higher or

lower range in the unidirectional or

bidirectional 2-range setting.

Totalizer control — Starts and stops the built-in

totalizer.

Fixed-value outputs —Outputs fixed-values for

current and pulse outputs.

Zero adjustment — Executes zero adjustment

(on-stream at zero flow rate).

Output signals

Current output:

4–20mAdc (load resistance 0 to 750

Ω)

Note: The current output cannot be used with

the PROFIBUS-PA ccommunication.

Digital outputs — One point (std.) and one more

point is optionally available as follows.

Digital output DO1 (std.):

Output type: Transistor open collector

Number of outputs: One point

Output capacity: 30Vdc, 200mA maximum

Digital output DO2 (opt.):

Output type: Solidstate relay output (non

polarity)

Number of outputs: One point

Output capacity: 150Vdc, 150mA maximum

or 150Vac (peak to peak), 100mA maximum

DO1 and DO2 functions — One of the following

functions can be assigned to DO1 (std.) and/or

DO2 (opt.)
Pulse output (available only for DO1,DO2)

Pulse rate: 3.6 to 36,000,000 pulses/hr (DO1)

3.6 to 360,000 pulses/hr (DO2)

(Over 3,600,000 pulses/hr, auto-setting)

Pulse width: 0.5 to 500ms (but less than half of

the period for 100% flow rate)

Note: The same and simultaneous pulse is not

available between DO1 and DO2.)

Multi-range selection outputs (Note 1)
High, High high, Low, and/or Low low alarm

outputs (Note 2)

Empty pipe alarm output
Digital Output Active Status (DO1 and DO2)

(Note 2)

Preset count output

Converter failure alarm output

Note 1: Two outputs (DO1 and DO2) are needed

for 4-range switching and forward/reverse

2-range switching.

Note 2: Normal Open (default set) or Normal

Close is selected for alarm outputs when

programming.

When power failure occurs, unit will be fault to
Normal Open.

Communications output:

• HART (std.) Digital signal is superimposed

on 4–20mAdc current signal as follows:

Conforms to HART protocol

Load resistance: 240 to 750

Load capacitance: 0.25µF maximum

Load inductance: 4mH maximum

PROFIBUS (opt.)

Protocol : PROFIBUS-PA
Baud rate : 31.25kbps
Bus voltage : 9-30VDC

Consumption electric current of bus:less than 16mA

Manufacture Ident-No. : 093B

HEX

Standard Ident-No. : 9740

HEX

Slave address : 0-126 (Default address is 126)
Profile : Profile Ver.3.01 for Process Control

Devices

Function blocks : AI(Flow)×1 , Totalizer×1

LCD display: Full dot-matrix 128×128 dot LCD

display (back-light provided)

The data on the LCD inside the converter can

rotate to 90, 180, and 270 degrees by a software,

without rotating the indicator itself. (Combined

type only)

Parameter settings — Parameters can be set as

follows:

• IR Switches: Three key switches are provided to

set configuration parameters.

• Digital communication: The AF900 hand-held

terminal or PROFIBUS is needed to set

parameters.

•Zero adjustment: Zero point adjustment can be

started by pressing the switch in the converter.

•Damping: 0.5 to 60 seconds (selectable in one

second increments)

“Field re-verification”

Mag-Prover

– Toshiba’s

Zero span calibration tool allows unit to be
re-calibrated and verified using an internal
software program. (For more information contact
Toshiba International Corp.)

Conditions when power fails:

Parameter setting values are stored in non-volatile

memory and the values will be restored when the

power returns to normal condition. The outputs

and display will remain as follows when power

fails.

• Current output: 0mAdc

Advertising