Caution, 4 grounding – Toshiba LF414 User Manual

Page 29

Advertising
background image


- 28 -

6F8A0870

4.4 Grounding

CAUTION

„ Do not wire cables and replace parts

when power is supplied.

„ Do not work on piping and wiring with wet hands.

DON’T

Wiring work and replacing
parts in the power-on state
may cause electric shock.

DON’T

Wet hands may result in electric shock.

(1) Grounding of the LF414/LF600F, LF414/LF610F type

Ground as shown in Figure 4.10. Make the grounding wire as short as possible. Use grounding
wire material of IV wire 5.5mm

2

or more. Do not share a grounding wire with other

instruments where grounding current may flow. (An independent grounding is preferable.)







Figure 4.10 Grounding the LF414/LF600F, LF414/LF610F Type

Grounding terminal

• If the piping material is conductive, connect
the grounding wires to the both ends of the
piping flange.

• If the piping material is non-conductive, perform
grounding (grounding resistance 100Ω or less).

Conductive material pipe
Example: Metal, etc.

Piping of non-conductive material
Example: Resin or metal piping whose
inside is resin-lined, etc.

Integral

Grounding wire

grounding resistance
100Ω or less

Advertising