2 troubleshooting, 1 flow rate is not indicated – Toshiba LF414 User Manual

Page 41

Advertising
background image


- 40 -

6F8A0870

7.2 Troubleshooting

If a problem occurs while using the LF414/LF600F, LF414/LF610F and LF414, follow the flowcharts
described below.

You may find a way to solve the problem. The flowcharts are based on three symptoms (1) to (3).

If you cannot solve the problem, contact your nearest Toshiba representative.

7.2.1 Flow rate is not indicated





















YES

Are power supplies correct
for each device?

Use the correct power supply for each
device.

Are power and I/O cables
installed correctly?

Install the cables correctly.
Refer

to

Chapter

5,

“Wiring.”

Is the fuse burnt?

Replace the fuse with a new one.

Is the flow range correctly set?

Set correctly.
Refer to combined converter's manual.

Is the flow direction match
with the arrow mark on the
detector?

Install the detector correctly.
Refer

to

Chapter

4,

“Installation”

Contact your nearest Toshiba
representative.

NO

NO

YES

NO

NO

START

YES

YES

NO

YES