Accessing information on the device

The display enables you to view the different information on the settings of the GoodKnight 420G.
(Refer to reference 1 in Figure 2, page 5).
The setting information is accessible whether the device is on standby or in operation. For this, just
press the Information Access button for one second. (See reference 2 in Figure 2, page 5). With each
additional key press, you access further data.

NOTE

In the following screens, the device is shown in operation mode and all the values are given as a guide
only.

10.0

cmH

2

O

P

Press for one

second on

Information

Access button

The pressure level set by the doctor (equivalent to the
prescribed pressure when the device is running).

5

min

Press

Information

Access button

The ramp time set by the doctor (period of time over which
the pressure gradually increases from the starting ramp
pressure to the pressure level).

4.0

cmH

2

O

P

Press

Information

Access button

The starting ramp pressure (pressure level at the
beginning of ramp); this parameter is adjustable and visible
only if the ramp time is other than 0.
The symbol used shows that the parameter is adjustable.

You can increase or reduce the value displayed by using the
ON/OFF or Ramp keys. For further details, refer to the
chapter “Using the ramp”. (See page 12).

114

h

Press

Information

Access button

The hour meter value (device operating time).

42

h

Press

Information

Access button

The compliance meter value (time during which the
patient has breathed with the device).

Id *.*

Press

Information

Access button


The GoodKnight 420G on-board software version.

Press the Information Access button again to exit the parameter setting menu.