Safety, Oxygen safety, Warning – Activeforever RemRest 901 CPAP Machine User Manual

Page 21

Advertising
background image

RemRest 21

Safety

ELECTROMAGNETIC
SUSCEPTIBILITY
This device meets the
requirements of IEC
60601-1-2 which describes
requirements for
electromagnetic
compatibility (EMC) of
medical devices. This
includes immunity to
radio frequency electric
fields and electrostatic
discharge, in addition to
the other applicable
requirements of the
standard. Compliance with
EMC standards does not
mean a device has total
immunity; certain devices
(cellular phones, pagers,
etc.) can interrupt
operation if they are used
near medical equipment.

WARNING
This unit is shipped with a
dust cover on the DB9
connector. This cap should
be left on except when
the connector is in use.

WARNING
This device is not suitable
for use in the presence of
a flammable anaesthetic
mixture with air or with
nitrous oxide.

OXYGEN SAFETY

Oxygen is a prescription drug and should

only be administered under the supervision

of a physician.

Oxygen supports combustion. To avoid

possible physical injury do not smoke or

place the device near any heat source while

using your CPAP with oxygen.

WARNING

If oxygen is used with this CPAP machine,

the oxygen flow must be turned off when

the CPAP machine is not operating.

Explanation of the warning: When the CPAP

device is not in operation and the oxygen

flow is left on, oxygen delivered into the

ventilator tubing may accumulate within the

CPAP machine enclosure. Oxygen

accumulated in the CPAP machine enclosure

will create a risk of fire. This warning applies

to most types of CPAP machines.

NOTE

At a fixed flow rate of supplemental oxygen

flow, the inhaled oxygen concentration will

vary, depending on the pressure settings,

patient breathing pattern, mask selection,

and leak flow rate. This warning applies to

most types of CPAP machines.