Activeforever RemRest 901 CPAP Machine User Manual

Page 11

Advertising
background image

RemRest 11

5. If the ramp function is not already set,

simply pushing the ramp button on once

the unit is running will not start the ramp

function. To initiate the ramp function:

• If the device motor is running, turn off

the device by pushing on/off button

(ramp light and on/off light turn off)

• Push the ramp button on (ramp light

turn on)

• Turn ON the device by pushing on/off

button (ramp light and on/off light on).

The CPAP will start at the prescribed low

setting and accelerate through the ramp

time to prescribed pressure setting.

6. To turn off the ramp function: While the

unit is running, push the ramp ON/OFF

button. The ramp light will go out and the

unit will go to the prescription pressure

setting. When the device is turned on

again, the RemRest will accelerate to the

prescription setting without the ramp.

7. Turn OFF the device (ramp light and

on/off light turn off) by pressing the

ON/OFF button.

RemRest Tier 2 & Tier 3

There are two options to provide power to

the CPAP: AC or DC power.

1. For AC Power, plug one end of the AC

power cord into the AC inlet plug located

on the back of the RemRest device. Plug

the other end into the wall socket. Go to

Step 3.

2. If using the DC power cord with auto

adapter for operation of the RemRest in a