18

Rev.

3

Installation

Bipolar High-Stability Supply

g. Connect a coaxial cable between the BNC connectors (22) on the

rear of the two Model 601 Energy Absorbers — no BNC terminators

should be used.

h. Connect Model 08150PS Master unit terminal block positions CS,

S-, and S+ to each corresponding Slave terminal (18). Connect

Master to first Slave, first Slave to second Slave, etc.

i. Connect two jumpers (17) from terminal block position S- to M- and

from S+ to M+ on the Master

1

power supply unit.

j. Connect the CT power supply to the

TRANSDUCER POWER

connector (5) on the rear of the Model 430 Programmer.

k. Connect the current transducer output to the current transducer

SIGNAL

connector (4) on the rear of the Model 430 Programmer.

l. Connect the special DB15 analog I/O cable from the

PROGRAM

OUT

connector (14) on the back of the Model 430 Programmer to

the DB15

ANALOG I/O

connectors on the rear of the Master (7) and

each Slave (20) power supply unit.

m. Install an instrumentation cable between the magnet support

stand top plate connector (8) and one of the

MAGNET STATION

connectors (11) on the rear of the Model 430 Programmer.

n. Optional: Install an instrumentation cable between one of the

LHe /

TEMP

connectors (10) on the rear of the Model 430 Programmer

and the Model 13x Liquid Helium Level Instrument and/or

temperature instrument (9). Refer to section A.2 on page 152.

o. Optional: Install an instrumentation cable between the

QUENCH

I/O

connector (13) on the rear of the Model 430 Programmer and

Aux connector

J2

on the rear panel of the Model 13x Liquid Helium

Level Instrument (12). Refer to section A.6.2 on page 157.

p. Connect each device line cord and Model 601 power adapter from

the respective device to the appropriate power receptacle.

q. Remote communications via Ethernet and/or RS-232 can be

accomplished by connecting suitable cabling to the Model 430

Programmer rear panel

ETHERNET

and/or

RS-232

connectors.

1. Do not add terminal block jumpers to the slave units. Only the single or master unit is jum-