4 setup, Indicating and adjustment elements, Setup – VEGA VEGATOR 537 Ex User Manual

Page 10

Advertising
background image

10

VEGATOR 537 Ex

23465-EN-030723

4 Setup

Indicating and adjustment elements

1

Connection plan

2

Lockable fixing screw

3

Test

1)

channel 1

4

LED level relay 1

5

LED-fault signal channel 1

6

Test key

1)

channel 2

7

LED level relay 2

1 2 3 4 5 6

1 2 3 4 5 6

on

2

1

1

2

3

4

5

6

7

8

9

11

13

14

15

10

1

2

on

off

on

off

VEGATOR 537 Ex

TEST

TEST

on

off

12

537

VEGATOR

8

LED fault signal channel 2

9

LED supply voltage

10 Fixing screw
11 Multipoint connector
12 Change-over switch min./max. control
13 DIL switch block for channel 1
14 DIL switch block for channel 2
15 Transparent cover

Setup

1)

Recurring test acc. to WHG
In accordance with the type approval acc. to WHG (Z-65.11-154, Z-65.11-285), the recurring test acc. to
WHG can be carried out by pressing the test key on the VEGATOR signal conditioning instrument. It is
neither necessary to remove the sensor nor trigger a response by filling the vessels. This applies to
VEGASWING 61Ex, 63Ex, 81AEx and 83AEx with two-wire oscillator SW 60ZEx and SW E82ZEx.