1.3 Technical data

Mechanical data

Transmitter materials, wetted parts

Process connection

stainless steel 1.4571

Diaphragm

stainless steel 1.4571, Hastelloy C276, Tantalum,
PTFE foil on 1.4571

Extension

stainless steel 1.4571

Transmitter materials, non-wetted parts

Transmitter housing

- Alu die casting, powder-coated

- stainless steel 1.4571

Materials of the processing unit

Housing

high resistance plastic PBT (Polyester)

Ground terminal

stainless steel 1.4305

Window of the indication module

Lexan

Weight

Transmitter

approx. 0.8 … 8 kg (depending on isolating
system)

Processing unit

approx. 480 g

Adjustment and indicating elements (in the processing unit)

Menu-driven adjustment with
additional functions
- adjustment elements

4 keys

- indicating elements

DOT-matrix display, 3 lines with 7 digits each

Version without adjustment module
- adjustment element

8-pole mini coding switch for adjustment of the

bus address

Indication module

LC-display with
- bar graph (20 segments)
- digital value (4 digits)
- tendency indicators for rising or falling values

Electrical data

Adjustment ranges

Zero

adjustable from -20 … +95 % of nominal range

Span

adjustable from 3.3 … 120 % of nominal range

Supply and signal circuit

Supply voltage

9 … 32 V DC
(delivered by the segment coupler)

Current consumption

10 mA ± 1 mA

Integration time

0 … 50 s

Rise time

70 ms (ti = 0 sec; 0 – 63 %)

Product description