9

1)

In compensated temperature range of 0°C … +80°C, reference temperature 20°C.

2)

Acc. to IEC 770, item 6.1.2 relating to the nominal range.

Product description

Influence of the ambient temperature (without considering the isolating systems)

Average temperature coefficient
of the zero signal

1)

- Turn Down 1 : 1

< 0.15 %/10 K with accuracy class 0.25
< 0.05 %/10 K with accuracy class 0.1

- Turn Down up to 1 : 5

typ. < 0.225 %/10 K with accuracy class 0.25
typ. < 0.075 %/10 K with accuracy class 0.1

- Turn Down up to 1 : 10

typ. < 0.3 %/10 K with accuracy class 0.25
typ. < 0.1 %/10 K with accuracy class 0.1

Long-term stability

Long-term drift of the zero signal

2)

< 0.1 % per 2 years

Other actuating variables

Calibration position

upright, diaphragm points downwards

Vibration resistance

mechanical vibrations with 4 g and
5 … 100 Hz, tested acc. to the regulations of
German Lloyd GL-characteristics 2

Operating conditions

Ambient conditions

Ambient temperature

-40°C … +85°C

- with indicating module

-20°C … +70°C

Storage and transport temperature

-40°C … +85°C

Product temperature, depending on
isolating liquid

Temperature
p

abs

>

1 bar

/

p

abs

< 1 bar

- silicone oil

-40°C … +150°C

/

-40°C … +130°C

- silicone oil and cooling element

-40°C … +200°C

/

-40°C … +130°C

- silicone oil and capillaries 1 m

-40°C … +200°C

/

-40°C … +130°C

- vegetable oil

-10°C … +130°C

/

-10°C … +130°C

- vegetable oil and cooling element

-10°C … +200°C

/

-10°C … +130°C

- high temperature oil a. cooling element -10°C … +400°C

/

-10°C … +200°C

- high temperature oil and capillaries 1 m -10°C … +400°C

/

-10°C … +200°C