7 maintenance and fault rectification, 1 maintenance, 2 remove interferences – VEGA VEGABAR 14 User Manual

Page 18: 7maintenance and fault rectification

Advertising
background image

7

Maintenance and fault rectification

7

.1 Maintenance

I

f the instrument is used properly, no special maintenance is required

in normal operation.

7

.2 Remove interferences

T

he operator of the system is responsible for taking suitable measures

to rectify faults.

VEGABAR

14

offers maximum reliability. Nevertheless, faults can

occur during operation. These may be caused by the following, e.g.:

l

S

ensor

l

P

rocess

l

V

oltage supply

l

S

ignal processing

T

he first measure to be taken is to check the output signal. In many

cases, the causes can be determined this way and the faults rectified.

S

hould these measures not be successful, please call in urgent cases

the VEGA service hotline under the phone no. +49 1805 858550.

T

he hotline is available to you 7 days a week round-the-clock. Since

we offer this service world-wide, the support is only available in the
E

nglish language. The service is free of charge, only the standard

telephone costs will be charged.

?

4 … 20

mA signal not stable

l

no atmospheric pressure compensation

à Check the pressure compensation in the plug or via the

capillaries

?

N

o 4 … 20 mA signal

l

C

onnection to voltage supply wrong

à Check connection according to chapter "Connection steps"

and if necessary, correct according to chapter "Wiring plan"

l

N

o voltage supply

à Check cables for breaks; repair if necessary

l

O

perating voltage too low or load resistance too high

à Check, adapt if necessary

Reaction when malfunc-
tions occur

Failure reasons

Fault rectification

24

hour service hotline

Checking the 4 … 20 mA
signal

18