VEGA VEGABAR 14 User Manual

Page 24

Advertising
background image

Ambient conditions

A

mbient temperature

-

V

ersion with plug connector

-20 … +85 °C (-4 … +185 °F)

-

V

ersion with cable outlet

-20 … +60 °C (-4 … +140 °F)

S

torage and transport temperature

-

V

ersion with plug connector

-40 … +100 °C (-40 … +212 °F)

-

V

ersion with cable outlet

-40 … +60 °C (-40 … +140 °F)

Process conditions

T

he specifications of the pressure stage and the product temperature are used as an overview.

T

he specifications of the type label are applicable.

P

ressure stage, process fitting

-

T

hread 316L

PN

60

-

T

hread PVDF

PN

10

P

roduct temperature depending on the measuring cell seal

-

FKM (VP2/A)

-20 … +100 °C (-4 … +212 °F)

-

EPDM (A+P 75.5/KW75F)

-40 … +100 °C (-40 … +212 °F)

V

ibration resistance

mechanical vibrations with 4 g and 5 … 100 Hz

7)

Electromechanical data

A

ngled plug connector

-

V

ersion

4-

pin according to ISO 4400

-

S

crew terminals for cable cross-section

up to

2.5

mm² (AWG 14)

-

C

able gland

M

16 (

for cable: ø 4.5 … 10 mm)

C

ircular plug connector

4-

pin with screwed connection M12 x 1

C

able outlet

-

L

ength

5

m (16.4 ft)

-

M

in. bending radius

25

mm (with 25 °C/77 °F)

-

D

iameter approx.

6

mm

Voltage supply

O

perating voltage

8 … 30 V DC

P

ermissible residual ripple

U

ss

<

1 V

L

oad

see diagram

7)

T

ested according to the guidelines of German Lloyd, GL directive 2.

24

VEGABAR

14

9 S

upplement

22441

-EN

-111006

Advertising