VEGABAR 21

Set-up

Adjustment by considering the actual
pressure (life-adjustment)

The life-adjustment comprises two steps:
1 Adjustment of zero
2 Adjustment of span

The appropriate output current is displayed
via the DOT-matrix.

1 Adjustment of zero
(e.g. process pressure zero or vessel
empty)

Set with the “+“ or “–“-key the cur-
rent value to 4,000 mA. Then push
the “OK“-key.

2 Adjustment of span
(e.g. process pressure or vessel level max.)

Set with the “+“ or “–“-key the cur-
rent value to 20,000 mA. Then push
the “OK“-key.

Note:
- A modification of zero does not influence

the adjusted span.

- For standard adjustment of zero/span “+“

and “–“ are pushed simultaneously. The
value jumps directly to 4 mA/20 mA.

- In case of a high turn down generally the

“+“ and “-“-key should be pushed simulta-
neously due to the resolution.

- When pushing the “+“ or “–“-keys individu-

ally, the output current remains at the last
value, only after storing with the “OK“-key it
takes the adjusted value.

- It is also possible to adjust currents for

part fillings or part pressures, e.g. 8 mA for
25 % and 16 mA for 75 %. Then VEGABAR
calculates automatically the current values
for 0 % or 100 % (only possible with
>3,3 %).

Adjustment without considering the
actual pressure (dry adjustment)

The adjustment without pressure comprises
four steps:
1 Enquiry of the unit in which the adjustment

should be carried out

2 Offset correction
3 Adjustment of zero
4 Adjustment of span

The offset correction (only with overpressure)
defines the reference basis for the measure-
ment. It can be carried out:
- before or after the adjustment of zero and

span

- before or after installation of VEGABAR.

VEGABAR must be unpressurised when the
offset correction is carried out!

The adjustment without pressure can be
carried out assembled or unassembled (e.g.
workroom). An actually available pressure
has no meaning for the adjustment.

1 Enquiry of the unit

Choose with the “+“ or “–“-key the
unit. Save the selected unit with the
“OK“-key.

2 Adjustment of zero

Adjust zero with the “+“ or “–“-key
and save with the “OK“-key.

4 Adjustment of span

Adjust span with the “+“ or “–“-key
and save with the “OK“-key.

Zero

4,000

mA

Span

20,000

mA

Zero

0,0

mbar

Span

100,0

mbar

Adjust

without

press.

Unit

0,0

mbar

psi
kPa