VEGA VEGAPULS 45 4 … 20 mA; HART compact sensor User Manual

Page 55

Advertising
background image

VEGAPULS 42, 44 and 45 – 4 … 20 mA

55

Accuracy

1)

(typical values under reference conditions, all statements relate to the nominal measuring
range)

Characteristics

linear

Resolution, general

max. 1 mm

Resolution of the output signal

1.6 µA or 0.01 %

Accuracy

see diagram

Measurement characteristics

1)

(typical values under reference conditions, all statements relate to the nominal measuring
range)

Min. span between

full and empty

> 10 mm (recommended > 50 mm)

Frequency

26 GHz technology

Intervals

- two-wire sensor

0.6 s

- four-wire sensor

0.5 s

Beam angle (at -3 dB)

- VEGAPULS 42

22°

optional

18°, 10° and 8° when using large coupling horns

deviating from the nominal dimension

- VEGAPULS 44

DN 50, ANSI 2“

18°

DN 80, ANSI 3“

10°

DN 100, ANSI 4“

DN 150, ANSI 6“

- VEGAPULS 45

not applicable, as the meas. tube acts as
waveguide

Adjustment time (response time)

2)

> 1 s (dependent on the parameter adjustment)

Influence of the process temperature

cannot be measured at 0 bar;
at 5 bar 0.004 %/10 °K;
at 40 bar 0.03 %/10 °K

Influence of the process pressure

0.0265 %/bar

Adjustment time

2)

> 1 s (dependent on the parameter adjustment)

Emitted radar power (average)

0.717 µW

Received average emitted power

3)

- distance 1 m

0.4 … 3.2 nW per cm² (0.4 … 3.2 x 10

-9

W/cm²)

Technical data

1)

Similar to DIN 16 086, reference conditions acc. to IEC 770, e.g.
temperature 15 °C … 35 °C; moisture 45 % … 75 %; pressure 860 mbar … 1060 mbar

2)

The adjustment time (also actuating time, response time or adjustment period) is the time required by the
sensor to output the correct level (with max. 10% deviation) after a sudden level change.

3)

Average emitted power (electromagnetic energy) received by a body per cm² directly in front of the antenna.
The received emitted power depends on the antenna version and the distance.

0,5 m

30 m

10 mm

3 mm

-3 mm

-10 mm

Advertising