Instrument modification

Installation procedure:

• Place the new form seal (5) over the meas-

uring cell (the conical end must point to the
process connection!).

• Carefully insert pressure transmitter into

the process connection by turning clock-
wise.

• Tighten hexagon pressure screw (4) (width

across flats, 46 mm) (45 Nm).

• Turn the housing (1) of the pressure

transmitter to the original position.

The exchange of the form seal is now fin-
ished; the removed seal must be properly
disposed of.

1 Pressure transmitter
2 Headless screw (prevents overwinding and must not be

removed)

3 Adapter
4 Hexagon screw
5 Form seal
6 Process connection