2 mounting steps, 3 mounting steps external housing – VEGA VEGABAR 64 Profibus PA User Manual

Page 13

Advertising
background image

1

2

Fig. 4: Temperature ranges
1

Process temperature

2

Ambient temperature

4.2 Mounting steps

For mounting VEGABAR 64, a welded socket is required. You can find
these components in the supplementary instructions manual "Welded
socket and seals
".

Seal the thread with teflon, hemp or a similar resistant seal material on
the process fitting thread 1½ NPT.

à Screw VEGABAR 64 into the welded socket. Tighten the hexagon

on the process fitting with a suitable wrench. Wrench size, see
chapter "Dimensions".

Warning:
The housing must not be used to screw the instrument in! Applying
tightening force can damage internal parts of the housing.

Seal the flange connections according to DIN/ANSI with a suitable,
resistant seal and mount VEGABAR 64 with suitable screws.

Use the seal suitable for the respective process fitting. You can find
the components in the supplementary instructions manual "Welded
socket and seals
".

4.3 Mounting steps external housing

1

Mark the holes according to the following drilling template

2

Depending on the mounting surface, fasten the wall mounting
plate with 4 screws

Welding the socket

Sealing/Screwing in
threaded versions

Sealing/Screwing in
flange versions

Sealing/Screwing in hy-
gienic fittings

Wall mounting

VEGABAR

64 • Profibus PA

13

4 Mounting

28252

-

EN

-081209

Advertising