10 supplement, 1 technical data – VEGA VEGASON 63 4 … 20 mA_HART - four-wire User Manual
Page 40
40
10 Supplement
VEGASON 63 • 4 … 20 mA/HART - four-wire
28780-EN-130417
10 Supplement
10.1 Technical data
General data
Materials, wetted parts
Ʋ Process fitting
UP (glass-fibre reinforced polyester resin)
Ʋ Transducer diaphragm
316Ti
Ʋ Seal transducer diaphragm/process
fitting
EPDM
Materials, non-wetted parts
Ʋ Mounting strap
1.4301
Ʋ Compression flange
PPH, 316L
Ʋ Housing
Plastic PBT (polyester), Alu die-casting powder-coated,
316L
Ʋ Seal between housing and housing
cover
NBR (stainless steel housing), silicone (Alu/plastic hous-
ing)
Ʋ Inspection window in housing cover
Polycarbonate
Ʋ Ground terminal
316Ti/316L
Weight
2.7 … 5.7 kg (6 … 12.6 lbs), depending on the process
fitting and housing
Output variable
Output signal
4 … 20 mA/HART (active)
HART output values
Ʋ HART value (Primary Value)
Distance to the level
Ʋ HART value (Secondary Value)
Temperature
Ʋ HART value (3rd Value)
Distance to the level - scaled
Resolution
1.6 µA
Failure signal current output (adjustable) mA-value unchanged 20.5 mA, 22 mA, < 3.6 mA
Current limitation
22 mA
Load
< 500 Ω
4)
Damping (63 % of the input variable)
0 … 999 s, adjustable
Met NAMUR recommendation
NE 43
Input variable
Measured variable
distance between lower edge of the transducer and
product surface
Measuring range
Ʋ Liquids
up to 15 m (49.21 ft)
Ʋ Bulk solids
up to 7 m (22.97 ft)
Dead band
0.6 m (1.969 ft)
4)
With inductive load ohmic share min. 25 Ω/mH.