10 supplement, 1 technical data – VEGA VEGASON 63 4 … 20 mA_HART - four-wire User Manual

Page 40

Advertising
background image

40

10 Supplement

VEGASON 63 • 4 … 20 mA/HART - four-wire

28780-EN-130417

10 Supplement

10.1 Technical data

General data
Materials, wetted parts

Ʋ Process fitting

UP (glass-fibre reinforced polyester resin)

Ʋ Transducer diaphragm

316Ti

Ʋ Seal transducer diaphragm/process

fitting

EPDM

Materials, non-wetted parts

Ʋ Mounting strap

1.4301

Ʋ Compression flange

PPH, 316L

Ʋ Housing

Plastic PBT (polyester), Alu die-casting powder-coated,

316L

Ʋ Seal between housing and housing

cover

NBR (stainless steel housing), silicone (Alu/plastic hous-

ing)

Ʋ Inspection window in housing cover

Polycarbonate

Ʋ Ground terminal

316Ti/316L

Weight

2.7 … 5.7 kg (6 … 12.6 lbs), depending on the process

fitting and housing

Output variable
Output signal

4 … 20 mA/HART (active)

HART output values

Ʋ HART value (Primary Value)

Distance to the level

Ʋ HART value (Secondary Value)

Temperature

Ʋ HART value (3rd Value)

Distance to the level - scaled

Resolution

1.6 µA

Failure signal current output (adjustable) mA-value unchanged 20.5 mA, 22 mA, < 3.6 mA
Current limitation

22 mA

Load

< 500 Ω

4)

Damping (63 % of the input variable)

0 … 999 s, adjustable

Met NAMUR recommendation

NE 43

Input variable
Measured variable

distance between lower edge of the transducer and

product surface

Measuring range

Ʋ Liquids

up to 15 m (49.21 ft)

Ʋ Bulk solids

up to 7 m (22.97 ft)

Dead band

0.6 m (1.969 ft)

4)

With inductive load ohmic share min. 25 Ω/mH.