Note:
S

ince with the function "Quick level change > 1 m/min." the generation

of an average value of the signal processing is considerably reduced,
false reflections by agitators or vessel installations can cause
measured value fluctuations. A false echo memory is thus recom-
mended.

T

his function enables reading out parameter adjustment data as well

as writing parameter adjustment data into the sensor via the indicating
and adjustment module. A description of the function is available in the
operating instructions manual "Indicating and adjustment module".

T

he following data are read out or written with this function:

l

M

easured value presentation

l

A

djustment

l

M

edium

l

V

essel form

l

D

amping

l

L

inearisation curve

l

S

ensor-TAG

l

D

isplayed value

l

D

isplay unit

l

S

caling

l

C

urrent output

l

U

nit of measurement

l

L

anguage

T

he following safety-relevant data are not read out or written:

l

HART

mode

l

PIN

C

opy sensor data

C

opy sensor data?

Basic adjustment
I

f the function "Reset" is carried out, the sensor resets the values of the

following menu items to the reset values (see chart):

1)

Function

Reset value

M

ax. adjustment

0

m(d)

M

in. adjustment

M

eas. range end in m(d)

2)

M

edium

L

iquid

V

essel form

not known

1)

S

ensor-specific basic adjustment.

2)

D

epending on the sensor type, see chapter "Technical data".

Copy sensor data

Reset