Align the sensor in such a way that the polarisation marking on the
sensor flange is in the same plane as the tube holes or the tube
connection openings (see illustration: "VEGAPULS in a bypass tube").

100%

0%

1

> 500 mm

Fig. 14: VEGAPULS 66 in a bypass tube. The polarisation marking on the
process fitting must be in one plane with the tube holes or the tube connection
openings.

1

Marking of the polarisation direction

When the sensor is mounted on a bypass tube, the distance from
VEGAPULS

66 to the upper tube connection should be approx.

500 mm (19.69 in) or more. In case of extremely rough tube inner
walls, you should use an inserted tube (tube in tube) or a radar sensor
with tube antenna.

Information:
With VEGAPULS 66 the polarisation level is always in the centre
between two flange holes.