3 operation – VEGA VEGAPULS 66 (≥ 2.0.0 - ≤ 3.8) standpipe ver. Profibus PA User Manual

Page 10

Advertising
background image

A special version of VEGAPULS 66 is available for each area of
application. The version with flange and standpipe is particularly
suitable for the measurement of solvents and liquid gases.

The standpipe antenna is also suitable for vessels with foam
generation or for measurement of products with low dielectric figures
(DK > 1.6).

Measurement in a standpipe is not recommended for very adhesive
products.

For pressures up to 160 bar (2320 psi) and temperatures up to 400 °C
(752 °F) there are graphite seals and ceramic antenna cones
available.

The antenna of the radar sensor emits short radar pulses with a
duration of approx. 1 ns. These pulses are reflected by the product
and received by the antenna as echoes. The running time of the radar
pulses from emission to reception is proportional to the distance and
hence to the level. The determined level is converted into an
appropriate output signal and outputted as measured value.

Power supply via the Profibus DP/PA segment coupler or VEGALOG
571 EP cards. A two-wire cable according to Profibus specification
serves as carrier of both power and digital data transmission for
multiple sensors. The instrument profile of VEGAPULS 66 corre-
sponds to profile specification version 3.0.

The GSD (instrument master files) and bitmap files necessary for
planning your Profibus-DP-(PA) communication network are available
from the download section on the VEGA homepage www.vega.com
under "Services - Downloads - Software - Profibus". There you can
also find the appropriate certificates. In a PDM environment, an EDD
(Electronic Device Description) is also required to enable the full range
of sensor functions (also available as a download).A CD with the
appropriate files can be ordered via e-mail under [email protected] or
by phone from one of the VEGA agencies under the order number
"

DRIVER.S".

The backlight of the indicating and adjustment module is powered by
the sensor. Prerequisite is a certain level of operating voltage.

The data for power supply are specified in chapter "Technical data".

The optional heating requires its own operating voltage. You can find
detailed information in the supplementary instructions manual "Heating
for indicating and adjustment module
". This function is generally not
available for approved instruments.