VEGA VEGACAL 64 For connection to a signal conditioning instrument User Manual

Page 28

Advertising
background image

28

9 Supplement

VEGACAL 64 • For connection to a signal conditioning instrument

30319-EN-130917

1

2

-50˚C

(-58˚F)

50˚C

(122˚F)

40˚C

(104˚F)

-40˚C

(-40˚F)

80˚C

(176˚F)

0˚C

(32˚F)

100˚C

(212˚F)

150˚C

(302˚F)

Fig. 14: Ambient temperature - Process temperature
1 Process temperature

2 Ambient temperature

Dielectric constant

≥ 1.5

Electromechanical data - version IP 66/IP 67 and IP 66/IP 68; 0.2 bar
Cable entry/plug

1)

Ʋ Single chamber housing

– 1 x cable gland M20 x 1.5 (cable: ø 5 … 9 mm), 1 x

blind plug M20 x 1.5

or:

– 1 x closing cap ½ NPT, 1 x blind plug ½ NPT

or:

– 1␣x plug (depending on the version), 1␣x blind stopper

M20␣x␣1.5

Spring-loaded terminals

for wire cross-section up to 2.5 mm² (AWG 14)

Electromechanical data - version IP 66/IP 68 (1 bar)
Cable entry

Ʋ Single chamber housing

– 1 x IP 68 cable gland M20 x 1.5; 1 x blind plug

M20 x 1.5

or:

– 1 x closing cap ½ NPT, 1 x blind plug ½ NPT

Voltage supply
Operating voltage

Ʋ Non-Ex instrument

12 … 36 V DC

Ʋ EEx-ia instrument

12 … 30 V DC

Ʋ EEx-d-ia instrument

18 … 36 V DC

1)

Depending on the version M12 x 1, according to ISO 4400, Harting, 7/8" FF.