VEGA VEGAFLEX 67 (-40…+150°C PFA insulated) 4 … 20 mA_HART two-wire User Manual

Page 45

Advertising
background image

-

R

od: ø 10 mm (0.394 in)

approx. 620 g/m (6.7 oz/ft)

-

C

able: ø 4 mm (0.157 in)

approx. 80 g/m (0.86 oz/ft)

-

G

ravity weight

325

g (11.5 oz)

I

nsulation thickness

-

R

od version

2

mm (0.08 in)

-

C

able version

0.75

mm (0.03 in)

O

n the gravity weight and beginning of the cable:

2

mm (0.08 in)

P

robe length L (from seal surface), cannot be shortened

-

R

od: ø 10 mm (0.394 in)

up to 4 m (13.12 ft)

-

T

rimming accuracy - rod

<

1

mm (0.039 in)

-

C

able: ø 4 mm (0.157 in)

up to 32 m (105 ft)

-

T

rimming accuracy - cable

±

0.05 %

L

ateral load - rod: ø 10 mm (0.394 in)

4 N

m (3 lbf ft)

M

ax. tensile load with cable: ø 4 mm

(0.157

in)

2 KN (450

lbf)

Input variable

M

easured variable

L

evel of liquids

M

in. dielectric figure (lower medium)

by 10 higher than the ε

r

of the upper medium

M

in. layer thickness (upper medium)

>

100

mm (3.937 in)

T

he measured value deviates from the actual level

until the min. layer thickness is reached.

M

ax. layer thickness (upper medium)

-

D

ielectric figure (upper medium) 10

6

m (19.69 ft)

-

D

ielectric figure (upper medium) 5

10

m (32.81 in)

-

D

ielectric figure (upper medium) 2

20

m (65.61 in)

M

in. dielectric figure (upper medium) with

rod, cable version

ε

r

>

1.6

VEGAFLEX

67 • 4 … 20

mA/HART - two-wire

45

10 S

upplement

32303

-

EN

-120329

Advertising