VEGA VEGAWAVE series 60 Electronics module User Manual

Page 7

Advertising
background image

7 Compare the new oscillator with the old one. The type label

of the oscillator must correspond to that of the old

oscillator. This applies particularly to instruments used in

hazardous areas.

8 Compare the settings of the two oscillators. Set the

adjustment elements of the new oscillator to the same

setting of the old one.

Information:

Make sure that the housing is not rotated during the electronics

exchange. Otherwise the plug may be in a different position

later.

9 Insert the oscillator carefully. Make sure that the plug is in

the correct position.

10 Insert the two hold screws and tighten them
11 Insert the wire ends into the open terminals according to

the wiring plan

12 Press the opening lever of the terminals downward, you will

hear the terminal spring closing

13 Check the hold of the wires in the terminals by lightly

pulling on them

14 Check the tightness of the cable entry. The seal ring must

completely encircle the cable.

15 Screw the housing cover back on
The electronics exchange is hence terminated.

As a rule, the exchange of the electronics must be

documented internally when used in Ex applications.

Electronics module - VEGAWAVE series 60

7

Mounting

32356

-EN

-060822

Advertising