VEGA VEGAWAVE 62 - NAMUR User Manual

Page 30

Advertising
background image

Sensor length (L) - suspension cable FEP
(-40 … +150 °C/-40 … +302 °F)

0.6 … 80 m (1.969 … 262.47 ft)

Output variable

Output

Two-wire NAMUR output

Current consumption

-

Falling characteristics (max.)

2.4 mA uncovered/≤ 1 mA covered

-

Rising characteristics (min.)

1 mA uncovered/≥ 2.4 mA covered

-

Fault message

1 mA

Necessary processing system

NAMUR processing system according to
IEC

60947-5-6 (EN 50227/DIN 19234)

Modes (NAMUR output adjustable to falling or rising characteristics)

-

Min.

rising characteristic curve (High current when
immersed)

-

Max.

falling characteristics (Low current when immersed)

Ambient conditions

Ambient temperature on the housing

-40 … +80 °C (-40 … +176 °F)

Storage and transport temperature

-40 … +80 °C (-40 … +176 °F)

Process conditions

Measured variable

Limit level of solids

Process pressure

-1 … 6 bar/-100 … 600 kPa (-14.5 … 87 psig)

Process temperature with PUR suspension
cable

-20 … +80 °C (-4 … +176 °F)

Process temperature with FEP suspension
cable

-40 … +150 °C (-40 … +302 °F)

Product density

-

Standard

>

0.02 g/cm³ (0.0007 lbs/in³)

-

adjustable

>

0.008 g/cm³ (0.0003 lbs/in³)

Granular size

max. 10 mm (0.4 in)

Electromechanical data - version IP 66/IP 67 and IP 66/IP 68; 0.2 bar

Cable entry/plug

1)

-

Single chamber housing

l

1 x cable gland M20 x 1.5 (cable: ø 5 … 9 mm),
1 x blind stopper M20 x 1.5

or:

l

1 x closing cap ½ NPT, 1 x blind plug ½ NPT

or:

l

1 x plug (depending on the version), 1 x blind
stopper M20 x 1.5

1)

Depending on the version M12 x 1, according to ISO 4400, Harting, 7/8" FF.

30

VEGAWAVE

62 • - NAMUR

9 Supplement

32254

-EN

-120418

Advertising