VEGA VEGAFLEX 61 Profibus PA User Manual

Page 51

Advertising
background image

L

3 mm

(0.118

")

0

-3 mm

(-0.118

")

15 mm

(0.591

")

20 mm

(0.787”)

0,08 m

(3.15

")

0,3 m

(11.811

")

20 m

(787.4

")

1

Fig. 28: Deviation VEGAFLEX 61 in cable version, probe length L > 20 m in water
1

Dead zone - no measurement possible in this area

L

Probe length

L

0

20 mm

(0.787

")

0,25 m

(9.843

")

5 mm

(0.197

")

-5 mm

(-0.197

")

15 mm

(0.591

"

)

0,3 m

(11.811

")

20 m

(787.4

")

0,15 m

(5.906

"

)

1

1

Fig. 29: Deviation VEGAFLEX 61 in cable version, probe length L > 20 m in oil
1

Dead zone - no measurement possible in this area

L

Probe length

Influence of the ambient temperature to the sensor electronics

Temperature drift

0.03 %/10 K relating to the max. measuring range
or max. 0.3 %

Temperature drift - Digital output

3 mm/10 K relating to the max. measuring range or
max. 10 mm

VEGAFLEX

61 • Profibus PA

51

10 Supplement

31835

-

EN

-100426