.1 General instructions

Use the recommended cables (see chapter "Connecting to
power supply
") and tighten the cable gland.

You can give your VEGAKON 66 additional protection against
moisture penetration by leading the connection cable down-
ward in front of the cable entry. Rain and condensation water
can thus drain off. This applies mainly to mounting outdoors, in
areas where moisture is expected (e.g. by cleaning processes)
or on cooled or heated vessels.

Fig. 3: Measures against moisture penetration

The process fitting must be sealed if there is gauge or low
pressure in the vessel. Before use, check if the seal material is
resistant against the measured product and the process
temperature.

The max. permissible pressure is stated in chapter "Technical
data
" in the "Supplement" or on the type label of the sensor.

4

.2 Mounting information

Due to agitators, equipment vibration or similar, the probe can
be subjected to strong lateral forces.

During operation, the probe must not touch any installations or
the vessel wall. If necessary, secure the end of the probe
(insulated). Provide a suitable insulated support directly above
the probe end.

Moisture

Pressure/Vacuum

Agitators