VEGA VEGACAL 69 Profibus PA User Manual

Page 38

Advertising
background image

38

10 Supplement

VEGACAL 69 • Profibus PA

31179-EN-130917

Accuracy (according to DIN EN 60770-1)
Reference conditions according to DIN EN 61298-1

Ʋ Temperature

+18 … +30 °C (+64 … +86 °F)

Ʋ Relative humidity

45 … 75 %

Ʋ Air pressure

860 … 1060 mbar/86 … 106 kPa (12.5 … 15.4 psig)

Temperature error

Ʋ < 120 pF

< 1 pF

Ʋ > 120 pF

1 % of the current measured value

Linearity error

< 0.25 % of the complete measuring range

Ambient conditions
Ambient, storage and transport tempera-

ture

-40 … +80 °C (-40 … +176 °F)

Process conditions
Process pressure

-1 … 2 bar/-100 … 200 kPa (-14.5 … 29 psig)

Process temperature

Ʋ Flange of PTFE

-40 … +100 °C (-40 … +212 °F)

Ʋ Flange of PP

0 … +60 °C (+32 … +140 °F)

45°C

100°C

1

-1

0

2

0°C

-40°C

1

2

Fig. 23: Process pressure - Process temperature (flange of PTFE)
1 Process pressure

2 Process temperature

2

1

45°C 60°C

1

-1

0

2

0°C

Fig. 24: Process pressure - Process temperature (flange of PP)
1 Process pressure

2 Process temperature

Advertising