VEGA VEGAFLEX 67 (-40…+150°C) 4 … 20 mA_HART four-wire User Manual
Page 14
Advertising
more than 1 m (3.281 in) length, you must secure the probe by
fastening a suitable isolated brace or guy directly above the end of the
rod.
I
f there is a danger of the probe touching the vessel wall during
operation due to product movements or agitators etc., the measuring
probe should be securely fixed.
I
n the gravity weight there is a thread (M12), e.g. for a ring bolt
(
optional) - (article no. 2.27424).
M
ake sure that the probe cable is not completely taut. Avoid tensile
loads on the cable.
A
void undefined cable-vessel connections, i.e. the connection must be
either grounded reliably or isolated reliably. Any deviation from this
requirement can lead to measurement errors.
Fixing
14
VEGAFLEX
67 • 4 … 20
mA/HART - four-wire
4 M
ounting
31854
-EN
-120329