5 connect the sensor to the external housing, 1 preparing the connection, 2 connection procedure – VEGA VEGACAL External housing User Manual

Page 10

Advertising
background image

10

5 Connect the sensor to the external housing

External housing • VEGACAP, VEGACAL

31085-EN-130916

5 Connect the sensor to the external

housing

5.1 Preparing the connection

Follow the instructions in the operating instructions manual of the

sensor.

5.2 Connection procedure

Note:

The metal sensor housing cannot be opened. The electrical connec-

tion must be carried out on the instrument housing.
1. Insert the cable end through the cable entry on the external hous-

ing (instrument housing)

Tip:

Only run the cable loosely. The connection cable can only be sepa-

rated on the external housing (instrument housing) - except BNC plug.

If the sensor should be dismounted, the connection cable must be

free to take out the sensor.

1 2

1 2

3

Fig. 5: Connection of the sensor in the socket housing
1 red - conductor

2 blue - screening

3 black - grounding

2. Connect the cables to the terminals in the socket housing.

The cables are numbered and cannot be interchanged. The

ground cable (black with ground marking) must be connected to

the ground screw.
Conductor (1) wire colour: red

Advertising