5 Connecting to power supply

VEGACAL 67 • Profibus PA

31759-EN-130917

but the protective effect against high frequency interference signals

remains.

5.2 Connection procedure

Proceed as follows:
1. Unscrew the housing cover
2. If a display and adjustment module is installed, remove it by turn-

ing it to the left.

3. Loosen compression nut of the cable entry
4. Remove approx. 10 cm (4 in) of the cable mantle, strip approx.

1 cm (0.4 in) of insulation from the ends of the individual wires

5. Insert the cable into the sensor through the cable entry
6. Lift the opening levers of the terminals with a screwdriver (see

following illustration)

7. Insert the wire ends into the open terminals according to the wir-

ing plan

Fig. 6: Connection steps 6 and 7

8. Press down the opening levers of the terminals, you will hear the

terminal spring closing

9. Check the hold of the wires in the terminals by lightly pulling on

them

10. Connect the screen to the internal ground terminal, connect the

outer ground terminal to potential equalisation

11. Tighten the compression nut of the cable entry. The seal ring must

completely encircle the cable

12. Screw the housing cover back on
The electrical connection is hence finished.