9 supplement, 1 technical data – VEGA VEGACAL 69 For connection to a signal conditioning instrument User Manual

Page 25

Advertising
background image

25

9 Supplement

VEGACAL 69 • For connection to a signal conditioning instrument

31181-EN-130917

9 Supplement

9.1 Technical data

General data
Material 316L corresponds to 1.4404 or 1.4435
Materials, wetted parts

Ʋ Process fitting - flange

PP or PTFE

Ʋ insulation (fully insulated)

FEP

Materials, non-wetted parts

Ʋ Probe - double rod fully insulated:

ø 14 mm (0.551 in)

316L

Ʋ Plastic housing

plastic PBT (Polyester)

Ʋ Aluminium die-casting housing

Aluminium die-casting AlSi10Mg, powder-coated - basis:

Polyester

Ʋ Stainless steel housing - precision

casting

316L

Ʋ Stainless steel housing, electropol-

ished

316L

Ʋ Seal between housing and housing

cover

NBR (stainless steel housing, precision casting), silicone

(aluminium/plastic housing; stainless steel housing,

electropolished)

Ʋ Ground terminal

316L

Process fittings

Ʋ Flanges

DIN from DN 50, ANSI from 2"

Weight

1)

Ʋ Instrument weight

0.8 … 4 kg (0.18 … 8.82 lbs)

Ʋ Rod weight: ø 14 mm (0.551 in)

2000 g/m (22 oz/ft)

Sensor length (L)

0.2 … 4 m (0.656 … 13.12 ft)

Max. lateral load

10 Nm (7.4 lbf ft)

Output variable
Output signal

in the range of 4 … 20 mA

Suitable signal conditioning instruments e. g. VEGAMET 381, 391, 513, 514, 515, 624
Fault message

> 22 mA

Current limitation

28 mA

Load

see load diagram under Power supply

Damping (63 % of the input variable)

0.1 s

Met NAMUR recommendation

NE 43

Input variable
Measured variable

Level of liquids preferably in non-conductive vessels

Measuring principle

phase-selective admittance processing (PSA)

1)

Flange weight not considered

Advertising