VEGA Replacement housing for plics instruments User Manual
Page 11
Advertising
5 Insert the oscillator carefully.
Information:
The electronics of most sensor types is connected via a plug.
Make sure that the plug is in the correct position. The notch
must be in position "18.00 h".
1
Fig. 7: Plug position
1
Notch
6 Screw in and tighten the two screws with a Phillips
screwdriver size 1.
7 Connect the connection cables acc. to the operating
instructions manual of the respective sensor
8 Screw the housing cover back on
The housing exchange is finished.
For Ex applications, the housing exchange must be docu-
mented in general, internally.
Replacement housing for plics instruments
11
Mounting
31150
-EN