VEGA Replacement housing for plics instruments User Manual

Page 9

Advertising
background image

1

Fig. 3: Double chamber housing

1

Position of the electronics compartment

Proceed as follows:
1 Switch off power supply
2 Remove the housing cover of the electronics compartment

as well as with Alu double chamber housing of the

connection compartment

3 Disconnect the connection cables acc. to the operating

instructions manual of the respective sensor

4 Loosen the two screws of the electronics module with the

Phillips head size 1

2

1

Fig. 4: Loosen the screws

1

Electronics module

2

Screws (2 pcs.)

5 Pull the electronics out by holding the opening levers.

Proceed as follows:
1 Loosen the screws of the check angle and the locking ring

with a screwdriver size 1

Remove the housing

Replacement housing for plics instruments

9

Mounting

31150

-EN

-050802

Advertising