3 Mounting

Antenna extension up to 450 °C • for VEGAPULS 62 and 68

38316-EN-131119

3 Mounting

3.1 Mounting preparations

When the antenna extension is shipped with the radar sensor, calibra-

tion is carried out beforehand in the factory. The radar sensor is thus

adapted to the antenna extension. To ensure correct functioning of the

measurement, an unambiguous allocation of the antenna extension to

the radar sensor is necessary. For this purpose the antenna parts are

marked with the serial number of the sensor.
Note:

An incorrect combination will compromise the accuracy. Make sure

that sensor and antenna extension belong together.

Electromagnetic radar signals are polarized, i.e. aligned with a certain

plane. With an angled antenna extension as well as versions with

rinsing air connection, the bending axis must be aligned with this po-

larisation. The procedure is described in chapters "Assembly steps".

In case of very long antenna extensions or heavy mechanical loads,

some kind of support is required.

1

Fig. 4: Absorption of mechanical loads by supports
1 Support

The following tools are required for mounting:

Hexagon socket wrench, size 4

3.2 Mounting instructions

Antenna extensions must be mounted not only with mechanical but

also microwave-specific considerations in mind. To ensure reliable

functioning of the sensors, gap-free mounting of the antenna exten-

sion sections in the correct position is necessary.

Assignment

Polarisation

Support

Tools

Gap-free mounting