VEGA VEGABAR 53 Profibus PA User Manual

Page 35

Advertising
background image

35

6 Set up with the display and adjustment module PLICSCOM

VEGABAR 53 • Profibus PA

36723-EN-130321

This function enables reading out parameter adjustment data as well

as writing parameter adjustment data into the sensor via the display

and adjustment module. A description of the function is available in

the operating instructions manual "Display and adjustment module".
The following data are read out or written with this function:

Measured value presentation

Adjustment

Damping

Linearization curve

Sensor-TAG

Displayed value

Scaling unit (Out-Scale unit)

Positions after the decimal point (scaled)

Scaling PA/Out-Scale 4 values

Unit of measurement

Language

The following safety-relevant data are not read out or written:

Sensor address

PIN

Application

Copy sensor data

Copy sensor data?

The reset function resets all parameters adjusted by the user to the

delivery status and the peak values to the actual values.

Reset

Select reset?

Reset
Basic adjustment
Peak value, measured value
Peak value, temperature

Basic adjustment

The "Reset" "Basic adjustment" resets the values of the following

menu items:

Menu section

Function

Reset value

Basic settings

Sensor address

126

Zero/Min. adjustment

Measuring range begin

Span/Max. adjustment

Measuring range end

Density

1 kg/l

Density unit

kg/l

Copy sensor data

Reset