6 Set up with the indicating and adjustment module

VEGAPULS 66 • Profibus PA

36521-EN-121011

Delivery status: Restoring the parameter settings at the time of

shipment from the factory incl. the order-specific settings. A cre-

ated false signal suppression, user-programmable linearization

curve, measured value memory as well as event memory will be

deleted.

Basic settings: Resetting of the parameter settings incl. special

and laboratory parameters to the default values of the respective

instrument. Any created false signal suppression, user program-

mable linearization curve, measured value memory as well as

event memory will be deleted.

Setup: Resetting of the parameter settings to the default values of

the respective instrument . Order-related settings remain but are

not taken over into the current parameters. User-generated false

signal suppression, user-programmed linearization curve, meas-

ured value memory, echo curve memory as well as event memory

remain untouched. The linearization is set to linear.

False signal suppression: Deleting a previously created false

signal suppression. The false signal suppression created in the

factory remains active.

Peak values measured value: Resetting of the measured min.

and max. distances to the actual measured value.

Select the requested reset function [->] and confirm with [OK].
The following table shows the default values of VEGAPULS 66:

Menu

Menu item

Default value

Setup

Measurement

loop name

Sensor

Medium

Liquid/Water
Bulk solids/Crushed stones, gravel

Applications

Storage tank
Silo

Vessel form

Vessel bottom, dished boiler end
Vessel top, dished boiler end

Vesell height/

Measuring range

Recommended measuring range, see

"Technical data" in the supplement

Min. adjustment

Recommended measuring range, see

"Technical data" in the supplement

Max. adjustment

0,000 m(d)

AI FB1 Channnel PV (lin. perc.)
AI FB1 scaling

unit

Height
%

AI FB1 scaling

0.00 lin %, 0.00 %
100.00 lin %, 100.00 %

AI FB1 damping

0 s

Lock operation

Released