11 Supplement

VEGAFLEX 86 • Profibus PA

44229-EN-130910

Max. deviation - Version with reference

distance

± 10 mm (0.39 in)

Deviating upper dead band - Version with

reference distance

450 mm (17.7 in)

L

0

-2 mm (-0.079

")

2 mm (0.079

")

-10 mm (-0.394

")

5 mm (0.197

")

-5 mm (-0.197

")

0,15 m (5.906

"

)

0,03 m (1.18

"

)

0,1 m

(3.937

")

1

Fig. 41: Deviation VEGAFLEX 86 in coax version in water
1 Dead band - no measurement possible in this area

L Probe length

L

0

-2 mm (-0.079

")

2 mm (0.079

")

-10 mm (-0.394

")

10 mm (0.394

")

0,1 m (3.937

"

)

0,05 m

(1.968

")

0,1 m

(3.937

")

0,25 m (9.842

"

)

1

1

Fig. 42: Deviation VEGAFLEX 86 in coax version in oil
1 Dead band - no measurement possible in this area

L Probe length

Repeatability

≤ ±1 mm

Variables influencing measurement accuracy
Temperature drift - Digital output

±3 mm/10 K relating to the max. measuring range or

max. 10 mm (0.394 in)

Additional deviation through electromag-

netic interference acc. to EN 61326

< ±10 mm (< ±0.394 in)

Influence of the superimposed gas and pressure to the accuracy
The propagation speed of the radar impulses in gas or vapour above the medium is reduced by

high pressure. This effect depends on the superimposed gas or vapour and is especially large at

low temperatures.