VEGA VEGAPULS 62 (≥ 2.0.0 - ≥ 4.0.0) Foundation Fieldbus Approval according to LPR radio standard User Manual

Page 71

Advertising
background image

-

Number of digits

5

-

Size of digits

W x H = 7 x 13 mm (sensors with software ≥ 4.0.0,
hardware ≥ 2.0.0)
W x H = 5 x 8 mm (sensors with software ≤ 3.99,
hardware < 2.0.0)

Adjustment elements

4 keys

Protection rating

-

unassembled

IP

20

-

mounted into the sensor without cover

IP

40

Materials

-

Housing

ABS

-

Inspection window

Polyester foil

Integrated clock

Date format

Day.Month.Year

Time format

12 h/24 h

Time zone Ex factory

CET

Measurement electronics temerature

Resolution

1 °C (1.8 °F)

Accuracy

±1 °C

(1.8 °F)

Voltage supply

Operating voltage

-

Non-Ex instrument

9 … 32 V DC

-

Non-Ex instrument

9 … 32 V DC

-

EEx-ia instrument - Power supply FISCO
model

9 … 17.5 V DC

-

EEx-ia instrument - Power supply EN-
TITY model

9 … 24 V DC

-

EEx-d instrument

14 … 32 V DC

Power supply by/max. number of sensors

-

H

1 power supply

max. 32 (max. 10 with Ex)

Electrical protective measures

Protection, depending on housing version

-

Plastic housing

IP

66/IP 67

-

Aluminium housing, stainless steel
housing - investment casting, stainless
steel housing - electro-polished

IP

66/IP 68 (0.2 bar)

10)

10)

The prerequisites for maintaining the protection rating are a suitable cable
as well as correct mounting.

VEGAPULS

62 • Approval according to LPR radio standard

71

11 Supplement

41721

-EN

-120301

Advertising