VEGA VEGAFLEX 83 Profibus PA PFA insulated User Manual

Page 37

Advertising
background image

37

6 Set up with the display and adjustment module

VEGAFLEX 83 • Profibus PA

44226-EN-130910

Note:

The selection of the application has a considerable influence on all

other menu items. Keep in mind that as you continue with the param-

eter adjustment, individual menu items are only optionally available.
You have the option of choosing the demonstration mode. This mode

is only suitable for test and demonstration purposes. In this mode, the

sensor ignores the parameters of the application and reacts immedi-

ately to each change.
You have to enter the tube inner diameter when measuring in a

bypass or standpipe.

In this menu item, you can define the type of medium (product).
This menu item is only available if you have selected level measure-

ment under the menu item "Application".

You can choose between the following medium types:

Dielectric con-

stant

Type of medium Examples

> 10

Water-based liq-

uids

Acids, alcalis, water

3 … 10

Chemical mix-

tures

Chlorobenzene, nitro lacquer, aniline,

isocyanate, chloroform

< 3

Hydrocarbons

Solvents, oils, liquid gas

This menu item is only available, if you have chosen interface meas-

urement under the menu item "Application". In this menu item you can

enter if there is a superimposed gas phase in your application.
Only set the function to "Yes", if the gas phase is permanently pre-

sent.

This menu item is only available if you have selected interface meas-

urement under the menu item "Application". In this menu item you can

choose the type of medium of the upper medium.

Setup - Application - Me-

dium, dielectric constant

Setup - Application - Gas

phase

Setup - Application - Di-

electric constant

Advertising