VEGA POINTRAC 31 Foundation Fieldbus User Manual

Page 43

Advertising
background image

43

11 Supplement

POINTRAC 31 • Foundation Fieldbus

41779-EN-131119

L

Fig. 17: Data of the input variable
L Measuring range (range in which the switching point must lie)

Measuring range

152 mm (6 in) or 304 mm (12 in)

Analogue input

Ʋ Input type

4 … 20 mA, passive

Ʋ Internal load

250 Ω

Switching input

Ʋ Type of input - Open Collector

10 mA

Ʋ Type of input - Relay contact

100 mA

Output variable
Output

Ʋ Signal

digital output signal, Foundation Fieldbus protocol

Ʋ Physical layer

according to IEC 61158-2

Damping (63 % of the input variable)

0 … 999 s, adjustable

Channel Numbers

Ʋ Channel 1

Process value (switching status)

Ʋ Channel 8

Electronics temperature

Ʋ Channel 9

Pulse rate

Transmission rate

31.25 Kbit/s

Current value

10 mA, ±0.5 mA

Resolution, digital

> 0.1 mm (0.004 in)

Relay output
Output

Relay output (SPDT), floating spdt

Switching voltage

Ʋ Min.

10 mV

Advertising