VEGA VEGABAR 87 Profibus PA - Operating Instructions User Manual

Page 34

Advertising
background image

34

6 Set up with the display and adjustment module

VEGABAR 87 • Profibus PA

45047-EN-131107

Save your settings with [OK] and move with [ESC] and [->] to the

next menu item.
After the position correction is carried out, the actual measured value

is corrected to 0. The corrective value appears with an inverse sign as

offset value in the display.
The position correction can be repeated as often as necessary. How-

ever, if the sum of the corrective values exceeds 20 % of the nominal

measuring range, then no position correction is possible.

VEGABAR 87 always measures pressure independently of the pro-

cess variable selected in the menu item "Application". To output the

selected process variable correctly, an allocation to 0 % and 100 % of

the output signal must be carried out (adjustment).
During adjustment, the pressure is entered e.g. for the level with full

and empty vessel, see following example:

2

1

100%

0%

2

m

(78.74

")

Fig. 28: Parameter adjustment example "Min./max. adjustment, level measure-

ment"
1 Min. level = 0 % corresponds to 0.0 mbar

2 Max. level = 100 % corresponds to 196.2 mbar

If these values are not known, an adjustment with levels of for exam-

ple 10 % and 90 % is also possible. By means of these settings, the

real filling height is then calculated.
The real product level during the adjustment is not important, because

the min./max. adjustment is always carried out without changing the

product level. These settings can be made ahead of time without the

instrument having to be installed.
Note:

If the adjustment ranges are exceeded, the entered value will not be

accepted. Editing can be interrupted with [ESC] or corrected to a

value within the adjustment ranges.

Parameterization example

Advertising