VEGA VEGABAR 87 Profibus PA - Operating Instructions User Manual

Page 63

Advertising
background image

63

11 Supplement

VEGABAR 87 • Profibus PA

45047-EN-131107

Adjustment elements

4 keys

Protection rating

Ʋ unassembled

IP 20

Ʋ mounted into the housing without

cover

IP 40

Materials

Ʋ Housing

ABS

Ʋ Inspection window

Polyester foil

Interface to the external display and adjustment unit
Data transmission

digital (I²C-Bus)

Configuration, connection cable

4-wire, screened

Cable length max.

Ʋ Sensors with signal output

4 … 20 mA, 4 … 20 mA/HART

50 m

Ʋ Sensors with signal output Profibus

PA, Foundation Fieldbus

25 m

Interface to the slave sensor
Data transmission

digital (I²C-Bus)

Configuration, connection cable

4-wire, screened

Cable length max.

25 m

Integrated clock
Date format

Day.Month.Year

Time format

12 h/24 h

Time zone Ex factory

CET

Rate deviation max.

10.5 min/year

Measurement electronics temerature
Resolution

1 °C (1.8 °F)

Accuracy

±1 °C (1.8 °F)

Voltage supply
Operating voltage

Ʋ Non-Ex, Ex-d instrument

9 … 32 V DC

Ʋ EEx-ia instrument - Power supply

FISCO model

9 … 17.5 V DC

Ʋ EEx-ia instrument - Power supply

ENTITY model

9 … 24 V DC

Number of sensors per DP/PA segment coupler, max.

Ʋ Non-Ex

32

Ʋ Ex

10