VEGA VEGABAR 81 4 … 20 mA_HART - Operating Instructions User Manual

Page 38

Advertising
background image

38

6 Set up with the display and adjustment module

VEGABAR 81 • 4 … 20 mA/HART

45018-EN-131010

If these values are not known, an adjustment with levels of for exam-

ple 10 % and 90 % is also possible. By means of these settings, the

real filling height is then calculated.
The real product level during this adjustment is not important, be-

cause the min./max. adjustment is always carried out without chang-

ing the product level. These settings can be made ahead of time

without the instrument having to be installed.
Note:

If the adjustment ranges are exceeded, the entered value will not be

accepted. Editing can be interrupted with [ESC] or corrected to a

value within the adjustment ranges.
For the other process variables such as e.g. process pressure, differ-

ential pressure or flow, the adjustment is performed in like manner.

Proceed as follows:
1. Select the menu item "Setup" with [->] and confirm with [OK].

Now select with [->] the menu item "zero adjustment" and confirm

with [OK].

2. Edit the mbar value with [OK] and set the cursor to the requested

position with [->].

3. Set the requested mbar value with [+] and store with [OK].
4. Move with [ESC] and [->] to the span adjustment
The zero adjustment is finished.
Information:

The zero adjustment shifts the value of the span adjustment. The

span, i.e. the difference between these values, however, remains

unchanged.
For an adjustment with pressure, simply enter the actual measured

value indicated at the bottom of the display.
If the adjustment ranges are exceeded, the message "Outside param-

eter limits" appears. The editing procedure can be aborted with [ESC]

or the displayed limit value can be accepted with [OK].

Proceed as follows:
1. Select with [->] the menu item "Span adjustment" and confirm

with [OK].

Setup - zero adjustment

Setup - span adjustment

Advertising