Specifications – Velleman DVM9912 User Manual

Page 10

Advertising
background image

DVM9912

VELLEMAN

10

8.

Specifications

Compliance

EN61010-1

Insulation

class 2, double insulation

Overvoltage capacity

CATIII 600V

Display

4000 counts LCD with function indication

Polarity

automatic, (-) negative polarity indication

Overrange

"OL" mark indication

Low Battery indication

"BAT" is displayed when the battery voltage drops below the
operating level

Measurement rate

2 times per second, nominal

Auto power off

meter automatically shuts down after approx. 15 minutes of
inactivity

Operating environment

0°C to 50°C (32°F to 122°F) at < 70% relative humidity

Storage temperature

-20°C to 60°C (-4°F to 140°F) at < 80% relative humidity

Use

inside use, max. height 2000m

Pollution degree

2

Power

1 x 9V battery, NEDA 1604, IEC 6F22

Dimensions

150 (H) x 70 (W) x 48 (D) mm

Weight

approx. 255g

Accuracy is given at 18°C to 28°C (65°F to 83°F), less than 70% RH

DC Voltage (Autoranging)


Range

Resolution

Accuracy

400.0mV

0.1mV

+0.5% of rdg + 2 dgts

4.000V

1mV

40.00V

10mV

400.0V

100mV

+1.2% of rdg + 2 dgts

600V

1V

+1.5% of rdg + 2 dgts

Input Impedance: 7.8MΩ

Maximum Input: 600VDC or 600VAC rms.


AC Voltage (Autoranging except 400mV)


Range

Resolution

Accuracy

400.0mV

0.1mV

+1.5%of rdg + 15 dgts

4.000V

1mV

+1.2% of rdg + 3 dgts

40.00V

10mV

400.0V

100mV

+1.5% of rdg + 3 dgts

600V

1V

+2.0% of rdg + 4 dgts

Input Impedance: 7.8MΩ

Frequency Range: 50 to 400Hz
Maximum Input: 600V DC or 600V ac rms.