Saved settings – Ocean Optics NeoFox User Manual

Page 43

Advertising
background image

3: Using NeoFox with NeoFox Viewer Software

013-20000-009-02-201203

33

2. Wait long enough for the temperature and tau values to settle – this amount of time will vary

based on the probe type and the number of averages parameter. You should wait at least one

3.

Single Point Reset button for at least five seconds. The Single Point Reset control button

is recessed in the center back of the NeoFox unit.

the temperature probe or the

ave selected for your device.

eans that all

minute.

Push the

After five seconds, NeoFox will use the following three values to perform the single point reset:

 Tau will be the current value of Tau measured from the probe.
 Temperature will be either the current temperature reading from

fixed temperature depending on which temperature source you h
If no temperature is selected, the SPR will not occur.

 Percent oxygen will be the value that was set during the setup phase in the lab.
Once the Single Point Reset has been performed, NeoFox saves your settings. This m
of the parameters listed in

Saved Settings

will be saved in their current states.

Saved Settings

d onboard the NeoFox in its non-volatile memory. When the settings

are saved, the saved values will be loaded when the device turns on:

n

tion Factor

ttings

e

alue

eset Settings

en

ture

The following parameters are save

Measurement Parameters:

 Duty Cycle Seconds o
 Duty Cycle Seconds off
 Number of Averages
 Oxygen Units
 Salinity Correc

Calibration Settings:

 Temperature Se

 Temperature Sourc
 Fixed Temperature V

 Calibration Method
 Sensor Type
 Single Point R

 Single Point – Tau
 Single Point – Oxyg
 Single Point – Tempera