Ocean Optics NeoFox User Manual

Page 52

Advertising
background image

4: Using NeoFox Sport

42

013-20000-009-02-201203

Menu Tree Description

Menu Item

Description

Sensor
Readout

Enables you to view the sensed values for Time, O2, temperature and Tau. The
values displayed refresh every time a new sample is uploaded from NeoFox with the
appropriate values.

Prs

Displays the atmospheric pressure measured by the on-board pressure sensor on the
NeoFox Sport.

O2

Displays the oxygen unit label and the current oxygen value.

Temp (°C)

Displays the algorithm temperature in degrees Celsius

Tau (µs)

Displays the value of tau (fluorescence lifetime)

Sensor Setup

Allows you set the parameters shown below.

Single Pnt
Reset

Allows you to set the following parameter settings to be used when a single point
reset is performed:

Mole %: Default = 20.90%; Range = 000.00 – 100.00%

Temperature: Default = algorithm temperature; Range = 000.00 – 200.00 °C

SPR – Pressure (kPa): Default = On-board pressure sensor

Tau: Default = current value of Tau; Range = 00.000 – 50.0000

A message displays informing you that the single point reset has completed.

NOTE:

NeoFox must be configured to Single Point mode through NeoFox Viewer.

Data Logging

Allows you to enable datalogging to an SD card, set the datalogging interval and see
the filename used for the stored data. Default = current datalog interval. Range =
1000 – 99,999,999 ms

Oxygen Units

Enables you to set the units in which the O2 data is displayed. Choose from %
PPO

2

%, ppm , torr and µmol/L, mole%

Temperature

Allows you to set the source of the temperature, either none, the sensor temperature
or a fixed temperature in degrees Celsius (Default = current value of fixed
temperature variable; Range = 0.0 – 200.0 °C).

Pressure

Allows you to set the source of the pressure: either none, the sensor pressure or a
fixed pressure in kPa (Default = current value of fixed pressure variable; Range = 0 –
200 kPa)

Averaging

Set the running average of the measurements. Default = current value of variable.
Range = 1– 300.

Duty Cycle

Set the duty cycle time in seconds. Range = 1 – 999. Set Off time to 0 to make it
always run.