Analytical Industries GPR-2500 A Oxygen Analyzer User Manual

Page 25

Advertising
background image



25

6. Press the ENTER key to calibrate or MENU key to abort and return to SAMPLING mode.

0.000

ZERO
CALIBRTION
ENTER TO CALIBRATE
MENU TO ABORT



7. Allow approximately 60 seconds for the calibration process while the processor determines whether the signal

output or reading has stabilized within 60% of the full scale low range.

8. Both the Zero Calibrate and Span Calibrate functions result in the following displays:



PASSED

CALIBRATION




OR


FAILED

CALIBRATION




Satisfying users that the zero offset is reasonably acceptable for their application can be accomplished much

quicker. Unless the zero gas is contaminated or there is a significant leak in the sample connections, the
transmitter should read less than 100 oxygen within 5 minutes after being placed on zero gas.
The maximum zero calibration adjustment permitted is 60% of the lowest full scale range available, which normally

is 1%. Thus the maximum zero calibration adjustment or zero offset is 0.6% oxygen. Accordingly, the transmitter’s

ZERO has not been adjusted prior to shipment because the factory conditions are different from the application

condition at the user’s installation.
Factory Default Zero:
The software will eliminate any previous zero calibration adjustment and display the actual the signal output of the

sensor at a specified oxygen concentration. For example, assuming a zero gas is introduced, the display will reflect

an oxygen reading representing basically the zero calibration adjustment as described above. This feature allows

the user to test the sensor’s signal output without removing it from the sensor housing.

MAIN MENU


AUTO SAMPLE
MANUAL SAMPLE

CALIBRATE



24.5 C 100 KPA



>>>

CALIBRATION


SPAN CALIBRATE
ZERO CALIBRATE
DEFAULT SPAN

DEFAULT ZERO

24.5 C 100 KPA