7 troubleshooting – Analytical Industries AII-2000 Palm O2 Oxygen Analyzer User Manual

Page 12

Advertising
background image

6.5 Oxygen Sensor Replacement - Optional Remote Sensor
The design of the electronics is intended for only the Analytical Industries Inc.
AII-11-75-PO2 or AII-11-75-PO2R Oxygen Sensors. Use of a different oxygen
sensor may result in an erroneous oxygen reading.


NEVER - Open the oxygen sensor or probe the
sensing surface, refer to Section 10 in the event
the sensor should leak and someone comes in

contact with the electrolyte from inside the sensor.

With this configuration the integral oxygen sensor is re-
placed by a connector module (P/N A-3654). The external
oxygen sensor (P/N AII-11-75-PO2R) is connected to the
module by a cable (P/N CABL-1009) with phone plug and
locking nut attached at both ends. The cable is coiled and
extends to 6 ft.

Procedure - Optional Remote Sensor
1. Unscrew the locking nut from the connector located
at the rear of the oxygen sensor.
2. Remove the new replacement sensor from its shipping
packaging.
3. Insert the phone plug into the connector at the rear
of the oxygen sensor and finger tighten the locking nut.
4. Allow the new replacement oxygen sensor to stabilize
for approximately 30 minutes in its new environment.
5. Calibrate, see section 4.4, after replacing the remote
oxygen sensor.

20

7 Troubleshooting


If the recommended corrective action does not resolve the problem return the
device to the factory for service.

21

Symptom

Corrective Action

Device appears to be physi-
cally damaged

Turn device ON – if it successful passes
calibration – proceed

No display when analyzer is
turned ON

Replace batteries
Check battery polarity
Check and/or clean battery contacts

‘LO’ displayed when analyzer is
turned ON or in use.

Replace battery and calibrate device

‘ERR’ displayed when analyzer
fails calibration

Replace oxygen sensor

No response to keypad

Replace battery

Cannot turn device OFF

Calibration routine in process – wait until
completed

Reading displayed by LCD
does not change when oxygen
level changes

Replace sensor

Reading does not stabilize or
fluctuates erratically

Relocate analyzer away source of RF or
electromagnetic radiation emissions.
Wait 5 minutes and repeat calibration
Replace sensor, repeat calibration

After calibration in 100% dry
oxygen, analyzer reading drifts
more than 2% over 8 hours

Check primary oxygen delivery device
Replace sensor that is nearing the end of
its useful life

Advertising