8 Specifications

22

8.1 Spare Parts & Optional Accessories












23

Accuracy:

< 2% of FS range under constant conditions

Analysis: 0-100%

oxygen

Alarms: Analyzer

none

Calibration:

Certified dry 100% oxygen or air after 8 hrs of use

Compensation: Temperature

Connections:

1x16mm thread or o-ring diverter

Controls:

Soft touch keypad for ON/OFF and CAL

Dimensions:

2.72” x 4.1” x 1.35”; weight 7 oz. (196 grams)

Display:

3 digit LCD 1.1” x .625”; resolution 0.1% O

2

Flow Sensitivity:

None between 0.2 to 10 liters per minute

Humidity: Non-condensing

0-95%

RH

Linearity: + 1% under constant conditions

Pressure:

Inlet – ambient or regulated; vent - atmospheric

Power:

(2) 1.5V AA alkaline batteries; 13,000 hrs of use

Response Time:

90% of final FS reading in 10 seconds

Sensitivity:

< 0.5% of FS range

Sensor:

AII-11-75-PO2; optional remote sensor AII-11-75-PO2R

Sensor Life:

32 months in air at 25ºC and 1 atmosphere

Storage Temp.:

-20º to 60ºC (-4ºF to 140ºF) on intermittent basis

Temp. Range:

5º to 45ºC (41ºF to 113ºF)

Warm-up Time:

None

Warranty:

12 months analyzer; 12 months sensor (any application)

Spare Parts:

AII-11-75-PO2 Oxygen Sensor

BATT-1008 Battery (2x) 1.5V AA Alkaline

P-1087 Instructions for Use

A-1162 PCB Assy Main

FITN-1112-1 Flow Diverter, O-ring Seal

Optional Accessories:

HRWR-1157 Screwdriver

HRWR-1158 Lanyard